SensorRAEElectrochemical Conditioning StationUser's GuideThe SensorRAE is a 4-socket electrochemical(EC)sensor conditioning station that keeps RAE SystemsEC sensors ready for immediate use while they arenot installed in a RAE Systems multi-sensor product.EC sensors must reach an electrically stable stateto provide accurate,reliable performance.While itRAESensorRAEtakes less than an hour for a CO sensor to stabilize,it.comtakes at least 24 hours for a NH,sensor.SensorRAEmaintains sensors in a ready-to-use state.(Referto RAE Systems Technical Note TN-114 for moreinformation.)FeaturesPractical:Users with just one multi-sensor detectorcan have a complete suite of RAE Systemselectrochemical sensors ready to go.Convenient:The SensorRAE holds four easy-to-plug-in sensors at a time and is suitable for both low biasand high bias (NH NO)sensors.Durable:The sensors are protected in a rugged,compact,waterproof Pelican TM case with atransparent lid.Low-Maintenance:Four AAA batteries powerthe SensorRAE for at least 12 months and theconditioning station also includes a battery test buttonwith an LED indicator.SpecificationsSize:5.4”×4.0”×2.1”(13.7×10.2×5.3cm)Weight without sensors:10 oz(285 g),with batteriesRA里Power:4 AAA alkaline batteriesRating:WaterproofS Y S T E M Sby HoneywellSupports the following RAE Systems Products:RAE Systems by HoneywellMultiRAE PlusContact InformationMultiRAE-IRMain Office 3775 N.First StSan Jose,CA95134,USAAreaRAE,AreaRAE Gamma,and AreaRAE IAQTelephone 408.952.8200SentryRAETol-Free877.723.2878VRAEFax408.952.8480QRAE PlusEmail RaeSales@raesystems.comIAQRAEWebsite www.raesystems.comTechnical Support 888.723.4800Use with Rae Systems'Electrochemical Sensors:Tech@raesystems.comCO,H,S,SO,NO,NO,Cl,HCN,PH and NH