Guangdong TechnionThe Life Sciences and Bio-Engineeringsrael Institute of TechnologyCore Facility广东以色到理工学航Operating Instructions-LyophilizerManufacturer:ZirbusModel:Vaco5Safety:1.The instrument has very cold surfaces.Always check the chamber temperature beforeloading samples.Use protective gloves when necessary to prevent cold injury.2.The chamber cover is heavy and high,ask for assistance or use a stool.3.Place your samples at the center of the chamber to prevent accidental sample loss whenplacing the chamber cover.4.Very high pressure difference!At the end of the drying process bring the system topressure equilibrium gradually.Rotate the connector valve and the aeration valve veryslowly.5.Solvents other than water are not allowed in the lyophilizer.Consult with the stuff.6.Inspect every glass flask for any cracks before use.7.Inform the Core Facility staff for any safety issue or misuse of the instrument.Note:Samples for freeze drying need to be pre-frozen at a temperature below-50C for at least 4hours.It is better to use a-80 C freezer.1.Turn on the Lyophilizer using the switch on the right side of the instrument.Chamber lidValveChamberConnectorChamber stageAbsorbing gauzeChamber baseLyophilizer main parts2.Wait~5 seconds until the main interface shows up.31/85/1812:37:15Device wo.:6Program selectUser:Total throughput time:00:00P2Total throughput time:00:08P3Total throughput time:eSystem interface3.Set the operating parameters.Up to 10 operating programs can be set.: